2 Monitoring or Collecting Vital Signs
27
Assignment with GuardianSoftware
To assign a Cableless Measurement Device to a patient in GuardianSoftware:
1 Select the patient on the Chalkboard.
2 On the Equipment List tab, select the Cableless Measurement Device on the Available
Equipment
list.
3 Click Use for Patient to assign the device to the patient.
Controls Available with a Patient Monitor
The controls available when the Cableless Measurement Device is assigned to a patient monitor are
described in the table below.
WARNING
If a patient being monitored by Cableless Measurement Devices moves out of range of the patient
monitor, the measurements are available on the Cableless Measurement Device only. They are not
transmitted to the patient monitor or the Information Center. If this occurs, the
No Host Monitoring
message is displayed on the measurement device. The measurement device will also sound the INOP
Action At the Cableless
Measurement
Device
At the Patient
Monitor
At the IIC
SpO
2
Start SpO
2
Yes Yes No
Change SpO
2
Mode Yes Yes No
Select SpO
2
Repetition Time Yes Yes No
Assign SpO
2
Pod Yes Yes No
Remove SpO
2
Pod Yes Yes No
Change Alarm Limits No Yes No
Place Device in Standby No Yes Yes
Alarm Silence No Yes Yes
Alarm Off/Pause No Yes Yes
NBP
Start/Stop/Stat NBP Yes Yes Yes
Change NBP Mode Yes Yes No
Change NBP Repetition Time Yes Yes No
Change Alarm Limits No Yes No
Assign NBP Pod Yes Yes No
Remove NBP Pod Yes Yes No
Place Device in Standby No Yes Yes
Alarm Silence No Yes Yes
Alarm Off/Pause No Yes Yes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue CL SpO2 and is the answer not in the manual?

Philips IntelliVue CL SpO2 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue CL SpO2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals