10 Specifications
76
Measurement Validation: The blood pressure measurements determined with this device comply
with the American National Standard for Electronic or Automated Sphygmomanometers (ANSI/
AAMI SP10:2002/(R)2008 + A1:2003/(R)2008) in relation to mean error and standard deviation,
when compared to auscultatory measurements in representative patient population. For the
auscultatory reference the 5th Korotkoff sound was used to determine the diastolic pressure.
Alarm Specifications
When the Cableless Measurement Devices are used alone, without an assignment to a monitor or
telemetry device, no physiological alarms will be generated.
When the Cableless Measurement Devices are assigned to a monitor or telemetry device, alarms may
be announced at the host monitor (for the telemetry device, at the Information Center). See the
respective Instructions for Use for detailed alarm specifications.
Alarm delay times are as specified in the Instructions for Use of your patient monitor or IntelliVue
Information Center plus 4 seconds.
NOTE
If the SRR connection between the Cableless Measurement Device and the patient monitor or
telemetry device is interrupted, the patient monitor or telemetry device will issue an INOP within
5 seconds.
Telemetry Device Battery Runtime Specifications
IntelliVue CL Charging Station Specifications
The IntelliVue CL Charging Station is intended for recharging and performing battery maintenance
procedures for the IntelliVue Cableless Measurement Devices. The IntelliVue CL Charging Station is
intended for use in medically used rooms (including use within the patient vicinity) which fulfill the
Initial Cuff Inflation Pressure
Adult: 165 ±15 mmHg
Pediatric: 130 ±15 mmHg
Venipuncture Pressure Range
Adult: 20 to 120 mmHg in steps of 5 mmHg
Pediatric: 20 to 80 mmHg in steps of 5 mmHg
Venipuncture Pressure Accuracy
±10 mmHg
Cuff size detection
INOP, if neonatal cuff size is detected
IntelliVue CL NBP Pod Performance Specifications
Operating Mode Battery Runtime
ECG Only Telemetry Device with CL SpO
2
/NBP Pod connected
Telemetry Device is host to CL SpO
2
/NBP Pod
20 hours
ECG/SpO
2
(continuous) Telemetry Device with internal SpO
2
running
in continuous mode and CL SpO
2
/NBP Pod connected
Telemetry Device is host to CL SpO
2
/NBP Pod
8 hours