10 Specifications
75
Battery
Integrated rechargeable Li-Ion battery with battery gauge and cycle counter
Runtime (fully charged battery)
Typical 4 days @ 2 measurements per hour
Typical 5 days @ 1 measurement every 2 hours
Battery Recharge Time
Maximum 2.5 hours
Short Range Radio Specifications
Type built in interface with integrated antenna
Technology IEEE 802.15.4
Frequency band 2.4 GHz ISM (2.400 - 2.483 GHz)
Modulation DSSS (O-QPSK)
Bandwidth 5 MHz
Effective radiated power (ERP) max. 0 dBm (1 mW)
Electronic Article Surveillance (EAS)
EAS tag inside the housing
Real Time Clock Accuracy
less than 5 seconds per day (typically) as long as device is in power state
"Device on" or "Device off". Automatically synchronized with assigned
host system.
Measurement Ranges
Adult:
Systolic: 30 to 270 mmHg (4.0 to 36.0 kPa)
Mean: 20 to 255 mmHg (2.5 to 34.0 kPa)
Diastolic: 10 to 245 mmHg (1.5 to 32.0 kPa)
Pediatric:
Systolic: 30 to 180 mmHg (4.0 to 24.0 kPa)
Mean: 20 to 160 mmHg (2.5 to 21.0 kPa)
Diastolic: 10 to 150 mmHg (1.5 to 20.0 kPa)
Pressure Transducer Accuracy
(0 to 300 mmHg)
±3 mmHg @ 15 to 25 °C
±(3 mmHg or 2% whichever is greater) @ 10 to 40°C
Blood Pressure Measurement
Accuracy
According to ANSI/AAMI SP10:2002/(R)2008 + A1:2003
8 mmHg standard deviation
±5 mmHg mean error
Pulse Rate Measurement Range
40 to 300 bpm
Pulse Rate Measurement Accuracy
40 - 100 bpm: ±5 bpm
101 - 200 bpm: ±5% of reading
201 - 300 bpm: ±10% of reading
(average over NBP measurement cycle)
Measurement Time
Auto/manual/sequence mode:
Typical 40 seconds @ >60 bpm and normal adult cuff
Maximum 180 seconds
STAT Mode:
Typical 30 seconds @ >60 bpm and normal adult cuff
Maximum 180 seconds
STAT Mode Cycle Time
5 minutes
IntelliVue CL NBP Pod Performance Specifications