Troubleshooting Running the Service Utilities
3-44 PageWriter TC 70 Cardiograph Service Manual
To access the Diagnostic Test screen
1
From the Main screen, touch the Setup button.
The Setup and Services main screen appears.
2 From the Setup and Service Utilities screen, touch the Diagnostic and Test Utilities
screen.
3 Use the Tab key on the keyboard to move the cursor to the Enter Password field.
4 Use the keyboard to enter the password (000000) for Diagnostic Tests and touch the
Submit button.
The Diagnostic Test screen appears.
Figure 3-6 Diagnostic Test screen
Flash Memory Test
This test verifies that the onboard Flash memory is working properly.
To test the onboard Flash memory
1
Touch the Flash Memory Test button.
A message lets you know if the onboard Flash memory passed the test.
2 Select Close.
If the onboard flash card fails the test, try the test again.