Troubleshooting Troubleshooting Cardiograph Issues
3-22 PageWriter TC 70 Cardiograph Service Manual
Compact Flash (CF) Card/USB Memory Stick Issues
Print Error:
Unknown Error
Application timing issue 1 End the patient session and restart a
new patient session.
2 Restart the cardiograph. See
“Restarting the Cardiograph” on
page 3-34.
Table 3-7 Printer Issues (continued)
Symptom Possible Cause & Investigation Step Solution
Table 3-8 Compact Flash Card / USB Memory Stick Issues
Symptom Possible Cause &
Investigation Step
Solution
Occasionally,
copying ECGs from
archive to a USB
Memory Stick fails.
USB memory stick
subsystem time-out
failure
1 Retry the transfer operation.
Write operations occasionally fail for USB
memory sticks.
2 If this condition reoccurs persistently, replace the
USB Memory Stick. Not all USB memory sticks
are supported. See “Ordering Options and
Upgrades” on page 5-10 for information on
ordering a Philips supported USB memory stick.
Maximum number
of ECGs stored
message
CF card file system
limitation
Delete files from the Archive.
Cannot store
additional ECGs
message
CF card file system
limitation
Delete unwanted files from the CF card
Cannot access
internal Compact
Flash memory
message
Compact Flash may be
corrupted.
1 Run the Internal Compact Flash test from the
Service Utilities on page 3-45.
2 If the test passes, retry the operation.
3 If the test fails, the Compact Flash card needs to be
replaced.
The cardiograph must be serviced to replace the
Compact Flash card. Contact the Philips Response
Center.