Troubleshooting Running the Service Utilities
3-36 PageWriter TC 70 Cardiograph Service Manual
Running the Service Utilities
The Service Utilities enable you to:
Obtain information about the cardiograph, such as software versions, installed options,
networking information, and battery status.
Verify and optimize cardiograph performance using the diagnostic tests and utilities
To run the Service Utilities
1
From the Main screen, touch the Setup button.
The Setup and Services main screen appears.
Figure 3-2 Setup and Services Main screen
2
Touch the button next to Service Utilities.
The Setup and Service Utilities screen appears.