Troubleshooting Cardiograph Issues Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-25
TraceMasterVue Issues
Table 3-9 TraceMaserVue Issues
Symptom Possible Cause &
Investigation Step
Possible Solutions
Error. The lead
system used to
generate this ECG is
not supported on
the cardiograph.
Please print ECG
directly from
TraceMaster.
message
ECG XML version is
incompatible
Print the ECG from TraceMasterVue.
Error. The ECG
originated from a
legacy TraceMaster
system and cannot
be printed. Please
print ECG directly
from TraceMaste
r.
message
ECG XML version is
incompatible
Print the ECG from TraceMasterVue.
The Patient ID field
must be enabled in
order to transfer
ECGs to
TraceMasterVue.
Touch the [ON]
button underneath
the “Enabled”
column to activate
the Patient ID field
.
message
You attempted to
transfer an ECG without
a Patient ID to
TraceMasterVue.
1 Follow the on screen instructions to enter a Patient
ID.
2 Retry the transfer operation.
Required patient
information is
missing from the
current ECG
message
You attempted to
transfer an ECG without
the required information
to TraceMasterVue.
Enter the required information and attempt the
operation again.