Installing the Software Upgrade Installing PageWriter TC 70 Software
PageWriter TC 70 Cardiograph Service Manual 6-3
Installing the Software Upgrade
Before You Begin
Save configuration data to a USB memory stick. Although data such as custom
configurations, archived ECGs, and orders are preserved during the software upgrade, it is
recommended that you back up this data to a USB memory stick before upgrading the
software.
Transmit archived ECGs to TraceMasterVue or a USB memory stick before upgrading the
software.
To install the software upgrade:
1
Ensure that the Patient Interface Module (PIM) is securely attached to the connector on
the rear of the cardiograph.
Figure 6-5 Connecting the PIM to the cardiograph
2
Insert the USB memory stick into the PC card slot.
3 Connect the cardiograph to AC power. Verify that the green LED on the front of the
cardiograph is lit.
WARNING Always connect the cardiograph to AC power when performing the software upgrade. If
the cardiograph is not connected to AC power when performing the software upgrade it
may be damaged.