Troubleshooting Cardiograph Issues Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-29
Remote Query Issues
The Order Search
Range cannot be left
empty. Please enter
a number between
0 and 99.
message
You attempted to search
for an order without
specifying the search
range.
Specify a search range and retry the operation.
Table 3-10 Orders Issues (continued)
Symptom Possible Cause &
Investigation Step
Possible Solutions
Table 3-11 Remote Query Issues
Symptom Possible Cause &
Investigation Step
Possible Solutions
Remote Query
could not be
completed
message
XML application was
not successfully created
due to a lack of system
resources.
1 Retry the operation.
2 Restart the cardiograph. See “Restarting the
Cardiograph” on page 3-34.
Remote Query get
list error
message
TraceMaster returned
the wrong ECG list.
1 Check the cardiograph’s TraceMaster
configuration settings.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
3 Retry the operation.
Remote Query get
ECG error
message
TraceMaster returned
the wrong ECG file.
1 Check the cardiograph’s TraceMaster
configuration settings.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
3 Check to see if the TraceMaster server is up and
running.
Remote transfer
error appears
message
Unknown transfer error
including no reply and
unexpected reply.
1 Check the cardiograph’s TraceMaster
configuration settings.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
3 Retry the operation.