Troubleshooting Troubleshooting Cardiograph Issues
3-30 PageWriter TC 70 Cardiograph Service Manual
No reply received
from server
message
The LAN
connection may
have been
disconnected during
the transfer
operation.
Server may be
down.
1 Check the LAN connection.
2 Check to see if the TraceMaster server is up and
running.
Unexpected reply
received from
server
message
Server may be down.
1 Check the LAN connection.
2 Check to see if the TraceMaster server is up and
running.
Remote transfer
schema error
message
The XML schema is not
supported for the
selected TraceMaster
server.
Check the TraceMaster Server Version in the
cardiograph’s TraceMaster configuration settings.
No matching ECG is
found on the
selected server
message
The Patient ID may
be incorrect.
The ECG you are
looking for does not
exist on the selected
server.
1 Check the Patient ID.
2 Check to see if there is an ECG with the same
Patient ID.
Table 3-11 Remote Query Issues (continued)
Symptom Possible Cause &
Investigation Step
Possible Solutions