PageWriter TC 70 Cardiograph Service Manual 2-19
Figure 2-5 Overall Sensitivity Test Results
8
If the results fall outside of the + 5% range, run the test again. If you get the same results,
try running the test on a different cardiograph. If the test fails on the other cardiograph, the
simulator may be out of calibration. If the test passes on the other cardiograph, it may
indicate a problem with your cardiograph. Contact the Philips Response Center for further
assistance.
Cardiograph and Accessory Disposal
When the cardiograph has reached the end of its product life, dispose of it according to local
ordinances. When any of the cardiograph accessories reach the end of their product life,
dispose of these items in accordance with manufacturer instructions and local ordinances.