Troubleshooting Troubleshooting Cardiograph Issues
3-16 PageWriter TC 70 Cardiograph Service Manual
Configuration Screen Issues
Table 3-6 Configuration Screen Issues
Symptom Possible Cause &
Investigation Step
Solution
The maximum
number of
TraceMaster
connections has
been configured.
Please edit or delete
an existing
TraceMaster
connection.
message
There are a limited
number of TraceMaster
connections that can be
configured on the
cardiograph.
Remove one of the TraceMaster connections.
The maximum
number of
OrderVue inboxes
has been created.
Edit or delete an
existing inbox.
message
There are a limited
number of OrderVue
inboxes that can be
configured on the
cardiograph.
Remove one of the OrderVue inboxes.
Invalid XML version
detected. Please try
again.
message
Invalid version detected
when loading
configuration file.
Reconfigure the cardiograph manually.
Invalid XML
platform detected.
Please try again.
message
Invalid platform
detected when loading
configuration file.
Reconfigure the cardiograph manually.
Invalid XML file
detected. Please try
again.
message
Invalid configuration
file detected when
loading configuration
file.
1 Ensure you are selecting a configuration file.
2 Reconfigure the cardiograph manually.
Failed to implement
custom settings.
The cardiograph
will automatically
restart and restore
custom settings.
message
Corrupted configuration
file detected when
loading configuration
file.
1 Cardiograph automatically restarts and loads
current configuration.
2 Reconfigure the cardiograph manually.
Duplicate Profile
name. Please enter
a unique name.
message
You attempted to save
an exam profile under
an existing name.
Choose another name and save the exam profile.