Troubleshooting Cardiograph Issues Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-15
Some ECGs
retrieved from
TraceMaster and
printed on the
cardiograph look
different from
TraceMaster prints or
XLi printouts.
Older or non-
PageWriter TC 70
source ECG files were
retrieved. ECG was not
created by a PageWriter
TC 70 cardiograph.
Inspect the printed report date and information
Retrieving and printing older ECG files that were not
sourced by the PageWriter TC 70 cardiograph can have
the following differences when printed:
Algorithm version appears as HPxxx on
TraceMaster and
PH on PageWriter TC 70.
Pacer tick marks are not present on PageWriter TC
70 printout.
Frank lead system generates 3 rhythm traces with
flat line and no lead label on PageWriter TC 70
printout.
Custom lead names do not appear on PageWriter
TC 70 printout.
PageWriter TC 70 prints full interpretations
regardless of original Xli ECG settings.
Error occurred
when deleting the
selected ECGs.
Please check the
Error List below
message
Internal Compact Flash
card read/write error
1 Retry the operation.
2 Contact the Philips Response Center.
Error occurred
when transferring
the selected ECGs.
Please check the
Error List below for
more details.
message
Transmission error 1 Retry the operation.
2 Contact the Philips Response Center.
Do you want to
overwrite the
existing patient
information with
the selected order
message
You attempted to use an
order to overwrite
patient information of
an archived ECG.
Check to make sure that you want to overwrite the
information. If so, confirm the operation. If not, cancel
the operation.
Table 3-5 Archive Screen Issues (continued)
Symptom Possible Cause &
Investigation Step
Solution