Running the Service Utilities Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-37
Using the About the Cardiograph Screen
The About the Cardiograph screen provides important information about the cardiograph that
can help in the troubleshooting process. This screen provides information on:
software revision levels, including application, kernel, and PIM software
PIM option installed, 12-lead or 16-lead
networking information such as MAC and IP address
storage information, such as total RAM, percentage of RAM used, Compact Flash and
USB memory stick usage
battery status, such as capacity, voltage, temperature, percent charged, and cycle count
voltage monitor information, such as AC/DC input, backlight, and PIM power
installed options
total pages printed
To access the About the Cardiograph screen
X From the Setup and Service Utilities screen, touch the About the Cardiograph tab.
Figure 3-3 About the Cardiograph screen
To refresh the screen
X Touch the Refresh button to get the latest information.