Cardiograph and PIM Cleaning Cardiograph Care and Maintenance
PageWriter TC 70 Cardiograph Service Manual 2-3
Cardiograph and PIM Cleaning
To clean the cardiograph and PIM:
1
Unplug the AC power cord.
2 Ensure that the AC power indicator light (next to power button) is not lit.
3 Wipe the external surfaces of the cardiograph and the PIM with a soft cloth dampened in
any of the approved cleaning solutions listed below.
CAUTION When cleaning, avoid the lead wire connectors and patient data cable connectors.
Approved Cleaning Solutions
Mild soap and water
Isopropyl alcohol (consisting of 70% solution in water)
Chlorine bleach (6% sodium hypo chlorite content), 3% solution in water
Quaternary ammonium compounds (21% quaternary ammonium content), for example,
Steris Coverage Plus NPD at a consistency of one half ounce per gallon of water (one part
Coverage Plus NPD to 255 parts water)
CAUTION Do not use strong solvents or abrasive cleaning materials.
Do not spill liquids on the surface of the cardiograph.
Do not use any of the following to clean the cardiograph:
Acetone
Iodine-based cleaners
Phenol-based cleaners
Ethylene oxide sterilization
The cardiograph or PIM should not be autoclaved, ultrasonically cleaned, or immersed.