Cardiograph Care and Maintenance
2-2 PageWriter TC 70 Cardiograph Service Manual
Table 2-1 Recommended Frequency of Maintenance Tasks
Component Recommended
Frequency
Maintenance Task and page number
Cardiograph and
Patient Interface
Module (PIM)
Cleaning
Weekly “Cardiograph and PIM Cleaning” on page 2-3
Patient Data Cable
and Lead Wire
Cleaning
Weekly “Patient Data Cable and Lead Wire Cleaning” on
page 2-4
Reusable
Electrode
Cleaning
After each
patient use
“Reusable Electrode Cleaning” on page 2-5
Print Head
Cleaning
When necessary
due to uneven
print quality
“Cleaning the Print Head” on page 2-6
Battery
Calibration
When necessary,
due to inaccurate
battery status
information
displayed on
cardiograph
“Battery Calibration” on page 2-14
Touch screen
calibration
When necessary,
due to decreased
touch screen
performance
“Touch Screen Calibration” on page 2-16
Touch screen
cleaning
Weekly “Touch Screen Cleaning” on page 2-16
Cardiograph
Overall Sensitivity
Test
Yearly “Cardiograph Overall Sensitivity Test” on
page 2-18