Cardiograph Care and Maintenance Patient Data Cable and Lead Wire Cleaning
2-4 PageWriter TC 70 Cardiograph Service Manual
Patient Data Cable and Lead Wire Cleaning
Approved Cleaning Solutions
Mild soap and water
Isopropyl alcohol (consisting of 70% solution in water)
Chlorine bleach (6% sodium hypo chlorite content), 3% solution in water
Quaternary ammonium compounds (21% quaternary ammonium content), for example,
Steris Coverage Plus NPD at a consistency of one half ounce per gallon of water (one part
Coverage Plus NPD to 255 parts water)
To clean the patient data cable and lead wires:
1
Dampen a soft cloth with soapy water or with one of the disinfectants or cleaning agents
listed in “Approved Cleaning Solutions” (above).
2 Wring excess moisture from the cloth before cleaning.
CAUTION Do not:
Autoclave the patient data cable or lead wires or use ultrasonic cleaners
Immerse
Use abrasive materials
Wet the connectors