Troubleshooting Cardiograph Issues Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-11
Archive Screen Issues
Table 3-5 Archive Screen Issues
Symptom Possible Cause &
Investigation Step
Solution
Command
timed-out
message
The TraceMaster ECG
Management system is
busy.
Retry TraceMaster operation.
Transfer operations use network and server resource,
which may occasionally be unable to service the
requests.
ECGs with
incomplete required
patient information
fields cannot be
transferred
message
The required
information for the
ECG has not been
entered.
Enter the required ECG information and try the archive
operation again.
Selected ECG(s)
exceeds available
space on
TraceMaster
Remote site
message
The TraceMaster
system is out of
available disk space.
Contact the TraceMaster system administrator and
report the error.

Table of Contents