Table 3-7 Printer Issues
Symptom Possible Cause & Investigation Step Solution
The message Busy
appears after you
have sent a job to the
printer.
A new print job has been sent to the
printer and the printer cannot accept a
new print job.
Restart the cardiograph. See “Restarting
the Cardiograph” on page 3-34.
The message
USB
appears when
attempting to print a
rhythm strip.
A new print job has been sent to the
printer and the printer cannot accept a
new print job.
Restart the cardiograph. See “Restarting
the Cardiograph” on page 3-34.
Printing is disabled
with one battery
only, please check
the battery
message
Cannot print with one battery Insert the second battery.
Low battery
warning: printing is
disabled. Connect
the cardiograph to
AC power and
recharge the
batteries
message
appears after you
have sent a job to the
printer.
No battery or dead battery Plug the cardiograph into AC power or
replace the battery.
Field
Programmable
Gate Array (FPGA)
communication is
incorrectly
configured. Please
contact the Philips
Response Center
for further
assistance.
message
Internal error Contact the Philips Response Center.
Field
Programmable
Gate Array (FPGA)
is not connected.
Please contact the
Philips Response
Center for further
assistance.
message.
Internal error Contact the Philips Response Center.