Troubleshooting Troubleshooting Cardiograph Issues
3-32 PageWriter TC 70 Cardiograph Service Manual
Log File and Custom Settings Issues
Table 3-13 Log File Issues
Symptom Possible Cause &
Investigation Step
Possible Solutions
Error accessing file
message
You can't copy files
from a USB memory
stick when attempting
to load a configuration
or log file.
1 Retry the operation
2 Make sure the file exists on the USB memory
stick.
Error damaged file
message
You can't open a file
due to a corrupted file.
Retry the operation
File already exists
message
You attempted to save a
log file to an existing
file.
Rename the log file and retry the operation.
Input error message Invalid path and file
name were specified.
Check the path and file name and ensure they are
correct.
No filename
specified
message
You did not specify a
file name in the Browse
dialog box.
Specify a file name in the Browse dialog box.
Invalid filename
specified
message
You did not specify a
file name in the Save
Logs dialog box.
Specify a file name in the Save Logs dialog box.
USB memory stick
is not detected.
Please ensure that
the USB memory
stick is fully inserted
into the USB
connector (right
side of cardiograph).
message
Custom settings or log
files cannot be saved
because there is no USB
memory stick in the
cardiograph’s USB
connector.
Insert a USB memory stick into the cardiograph’s USB
connector (right side of cardiograph as the device is
facing you).
The cardiograph
must restart in
order to load a new
custom settings file
message
A configuration file has
been loaded.
Restart the cardiograph to use the new configuration