Viewing and Saving Log Files Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-47
Saving Log Files
The following log files can be saved to a USB memory stick:
LOG1.txt
LOG2.txt
CFG.xml
These log files can be sent to the Philips Response Center and are useful in troubleshooting
cardiograph issues.
To save log files
1
Insert the USB memory stick into the cardiograph.
2 From the Setup and Services main screen, touch the button next to Configure
Cardiograph Default Settings
.
3 Touch the Save/Load Settings button on the Configuration Context Toolbar. The Save/
Load Settings
screen appears.
Figure 3-9 Save/Load Settings screen
4
Under Save Logs, make sure that USB memory stick is highlighted.
5
Touch the Save button.
The Save Logs screen appears.