Troubleshooting Cardiograph Issues Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-27
Orders Issues
Table 3-10 Orders Issues
Symptom Possible Cause &
Investigation Step
Possible Solutions
Command cannot
be completed
message
The TraceMasterVue
system is busy.
Retry TraceMaster operation.
Transfer operations use network and server resource,
which may occasionally be unable to service the
requests.
Command canceled
message
You canceled the action
or specified invalid
information.
Check the information specified and try again.
Internal error
message
The orders list does not
refresh.
1 Retry the operation.
2 Contact the Philips Response Center.
There is not enough
internal storage
available to save
new orders
message
Application limitation Delete unwanted orders.
There is not enough
internal storage
available to
download new
orders
message
Application limitation Delete unwanted orders.
Not all orders can
be downloaded
because the
Worklist is full
message
Application limitation Delete unwanted orders.
Orders cannot be
downloaded
because they have
been previously
archived with ECGs.
Please check ECGs
saved to the Main
Archive.
message
Order to be downloaded
is already used to
archive an ECG.
Transmit the ECGs to TraceMasterVue.