Troubleshooting Troubleshooting Cardiograph Issues
3-26 PageWriter TC 70 Cardiograph Service Manual
There are no
TraceMaster
connections
configured for use.
Please go to Setup
and add a
TraceMaster
connection.
message
You attempted to access
a TraceMasterVue
server and no
connection has been
configured.
1 Check the cardiograph’s TraceMaster
configuration settings and ensure a valid
TraceMasterVue server is configured.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
There is no
TraceMaster server
selected as the
default
configuration.
Please go to Setup
and select a default
TraceMaster
connection.
message
You did not configure a
default TraceMasterVue
server.
1 Check the cardiograph’s TraceMaster
configuration settings and ensure a valid
TraceMasterVue server is configured as the default
TraceMaster server.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
Invalid transfer
destination is
selected
message
You selected an invalid
TraceMasterVue server.
1 Check the cardiograph’s TraceMaster
configuration.
2 Check the cardiograph’s LAN/WLAN settings.
The XML version
associated with this
ECG is not
supported by
TraceMaster
message
ECG XML version is
incompatible
1 Check the cardiograph’s TraceMaster
configuration settings and ensure the correct
TraceMaster server version is selected.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
There is no
TraceMaster server
to delete
message
You attempted to delete
a TraceMasterVue
server without first
selecting one.
1 Select the TraceMasterVue server to delete.
2 Delete the TraceMasterVue server.
Missing network
settings
information. Please
enter correct and
complete network
settings.
message
You attempted to
configure a
TraceMasterVue
without specifying all of
the correct information.
1 Check the cardiograph’s TraceMaster
configuration settings and ensure that all of the
necessary information has been entered.
2 Check the cardiograph’s LAN/WLAN
configuration settings.
Table 3-9 TraceMaserVue Issues (continued)
Symptom Possible Cause &
Investigation Step
Possible Solutions