Troubleshooting Cardiograph Issues Troubleshooting
PageWriter TC 70 Cardiograph Service Manual 3-23
There are more
than 200 ECGs
stored on the USB
memory stick. Only
the first 200 ECGs
will display on the
cardiograph. Please
store no more than
200 ECGs on the
USB memory stick.
message
USB memory stick file
system limitation
Delete unwanted files from the USB memory stick.
An unknown error
occurred when
accessing the
storage capacity of
the selected archive
message
Abnormal error occurs
when getting storage
capacity of selected
storage media.
1 Retry the operation.
2 If this condition reoccurs persistently, replace the
USB Memory Stick. Not all USB memory sticks
are supported. See “Ordering Options and
Upgrades” on page 5-10 for information on
ordering a Philips supported USB memory stick.
The USB memory
stick is not
accessible. Please
ensure that the USB
memory stick is
fully inserted into
the USB connector
(right side of
cardiograph).
message
The USB memory
stick was not fully
inserted into the slot
on the cardiograph.
The USB memory
stick contains
unsupported
software (such as
SanDisk U3
Desktop software).
1 Verify that the USB memory stick is inserted
correctly.
Attempting to save ECGs to a USB memory stick
that is not accessible will result in an error
message.
2 Uninstall any software that is on the USB memory
stick and format the drive.
3 Retry the operation with fully functioning USB
memory stick that is fully inserted into the USB
connector on the cardiograph. Not all USB
memory sticks are supported. See “Ordering
Options and Upgrades” on page 5-10 for
information on ordering a Philips supported USB
memory stick.
The internal
CompactFlash (CF)
card cannot be
accessed. Please
contact the Philips
Response Center.
message.
Compact Flash may be
corrupted.
The cardiograph must be serviced to replace the
Compact Flash card. Contact the Philips Response
Center.
Table 3-8 Compact Flash Card / USB Memory Stick Issues (continued)
Symptom Possible Cause &
Investigation Step
Solution