Menu for Current Mode IG50
Current Mode IG50: monophase Current Mode IG50: biphase
Parameters for Current Mode IG50
Polarity switch to switch the polarity of the two electrodes ( the sign in the intensity dis-
play changes accordingly); can be adjusted separately for both channels.
Switching from monophase to biphase pulse emission; can be adjusted separately for
both channels.
Pulse shape Triangle
Stimulation fre-
quency
approx. 8 Hz
Pulse time 1 ms
Release time 20 ms
Threshold time 50 ms
Threshold inter-
val
70 ms
3.3.13 (FM) Frequency-Modulated Current
Definition Shiver-inducing frequency with automatically altered stimulation cycles.
Application
• Circulatory stimulation
• Relaxing muscle tensions
• Analgesia
• Sports physiotherapy with little sensitive strain
PHYSIOMED ELEKTROMEDIZIN AG
3 Overview of Current Modes
IONOSON-Basic 27