NOTICE
You can only change the pulse shape from triangle to square if the intensity in both
channels is zero. Square pulses make the current mode more intense.
3.3.15 (FaS) Faradic Surge Current
Definition Faradic surge current
Application
Treatment of muscle atrophies
Intentional exercises according to Förster
Electrogymnastics
Muscle training in sports
Coordination exercises
Menu for Current Mode FaS
Current Mode FaS: monophase Current Mode FaS: biphase
Parameters for Current Mode FaS
Polarity switch to switch the polarity of the two electrodes (the sign in the intensity dis-
play changes accordingly); can be adjusted separately for both channels.
Switching from monophase to biphase pulse emission; can be adjusted separately for
both channels.
Pulse duration T (1 - 60 s or continuously) and release time R (0 - 60 s or manual re-
lease key); can respectively be adjusted separately for both channels, refer to Dual
Channel Operation on page 11.
Ramp (Surge time) m
1
- m
4
for different slew rates; can respectively be adjusted separ-
3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG