Socket <11> on the front of the instrument and hand the manual release key to the patient.
(8) Slowly turn up the intensity of channel I with the Intensity Control Channel I <3>. The therapy
time elapses.
CAUTION
Make sure you do not touch the electrodes when the current is turned up!
(9) After treatment, an acoustic signal will be issued. The intensity is turned down to zero automatically.
Remove the electrodes and rinse the viscose covers thoroughly.
5.5.3 Pain Therapy, Hyperaemisation, Detonisation
Treatment with stochastic current
The following current modes can be used for these modes of therapy: G, GMC, DF, MF, CP, LP, UR, HV,
TENS, MENS, IG 30, IG 50, FM and STOCH.
You find a detailed overview about the parameters adjustable under:
Parameters for Current Mode G on page 16
Parameters for Current Mode GMC on page 17
Parameters for Current Type DF on page 17
Parameters for Current Mode MF on page 18
Parameters for Current Mode CP on page 19
Parameters for Current Mode LP on page 20
Parameters for Current Mode UR on page 21
Parameters for Current Mode HV on page 22
Parameters for Current Mode TENS on page 23
Parameters for Current Mode MENS on page 25
Parameters for Current Mode IG30 on page 26
Parameters for Current Mode IG50 on page 27
Parameters for Current Type FM on page 28
Parameters for Current Mode STOCH on page 29