3.3.16 (T/R) Pulses with Adjustable Parameters
Definition Pulses with adjustable parameters
Application
Treatment of peripheral paralysis
selective muscle stimulation
Menu for Current Mode T/R
Current Mode T/R: monophase Current Mode T/R: biphase
Parameters for Current Mode T/R
Polarity switch to switch the polarity of the two electrodes (the sign in the intensity dis-
play changes accordingly); can be adjusted separately for both channels.
Switching from monophase to biphase pulse emission; can be adjusted separately for
both channels.
Pulse shape (Square pulse r
1
, Triangle pulse d
1
or d
2
, exponential pulse e
1
); can be adjus-
ted separately for both channels.
Pulse length (100 µs - 1 s)
Release time (5 ms - 7 s or manual release key to release a single pulse)
Acoustic signal; makes pulses audible
3 Overview of Current Modes PHYSIOMED ELEKTROMEDIZIN AG