A.8 Technical Data
General Technical Data
Protection class acc. to IEC 60601-1/ VDE 750 Teil1 I, type: BF
CE characterization acc. to Council Directive concerning medical devices
(93/42 EEC)
Class acc. to Council Directive concerning medical
devices
IIa
Mains supply 100 ... 240 VAC ± 10%
Mains frequency 50 ... 60 Hz
Power consumption 140 VA
Current consumption 0,4 A ... 1,1 A
Ambient temperature (operation) + 10 °C ... + 40 °C
Dimensions (W x H x D) 31.5 x 10.4 x 33.5 cm
Weight 2.8 kg (battery-operated model) or. 2.6 kg (mains
operated model)
Output Data - Stimulation Current
Current Mode Output Data
GMC , MENS 1,000 µA
G 25 mA / 500 Ohm
HV and TENS 140 mA / 500 Ohm
DF , MF , CP , LP , UR , IG30 , IG50 , FM , STOCH , FaS , T/R 75 mA / 500 Ohm
Medium frequency current modes (IFM, AMF and KOTS) 100 mA / 500 Ohm
Output Data - Ultrasound - Small Transducer
Ultrasound Frequency 1 MHz 3 MHz
Average power density 3 W/cm² 1 W/cm²
Effective output 5 W 2.5 W
Maximum output 7.5 W 2.5 W
ERA 2.5 cm² 2,5 cm²
BNR <= 6 <= 6
Bundle type Diverging Diverging
Output Data - Ultrasound - Large Transducer
Ultrasound Frequency 1 MHz 3 MHz
Average power density 3 W/cm² 1 W/cm²
Effective output 10 W 5 W
Maximum output 15 W 5 W
ERA 5 cm² 5 cm²
PHYSIOMED ELEKTROMEDIZIN AG
Annex
IONOSON-Basic 67