3.4 Medium Frequency Current Modes
3.4.1 (IFM) Modulated Interference Current
Definition Medium frequency alternating current that features a outstanding compatibility and
significant depth effect.
Application
Pain therapy
Circulatory stimulation
Detonisation
Menu for Current Mode IFM
Current Mode IFM: Sinusoidal Waves
Parameters for Current Mode IFM
Vector function (on/off); varies the intensity of the two channels and thus
changes the effective field of the current
Frequency cycle (1 s / 30 s / 1 s; 1 s / 1 s / 1 s; 3 s / 3 s / 3 s; 6 s / 6 s / 6 s; 12 s
/ 12 s / 12 s; Time upper frequency / Changing time / Time lower frequency)
Basic frequency (2.0 - 9.5 kHz). A lower frequency enhances the stimulation
intensity, a higher frequency lowers it.
Frequency bands (1 - 25 Hz; 1- 50 Hz; 1 - 250 Hz; 100 - 250 Hz; program-
mable freely 1 - 250 Hz; or SELECTIVELY 1 - 250 Hz); the frequency currently
active is displayed in the frequency field next to the f symbol
Pulse shape Sinusoidal waves