3.3.14 (STOCH) Stochastic Current
Definition Stimulation current with random pulse triggering
Application
Circulatory stimulation
Analgesia with minimal stimulus adaptation
Menu for Current Mode STOCH
Current Mode STOCH: monophase Current Mode STOCH: biphase
Parameters for Current Mode STOCH
Polarity switch to switch the polarity of the two electrodes (the sign in the intensity dis-
play changes accordingly); can be adjusted separately for both channels.
Switching from monophase to biphase pulse emission; can be adjusted separately for
both channels.
Pulse shape (triangular pulse, square pulse); can be adjusted separately for both chan-
nels
Stimulation fre-
quency
10 - 100 Hz (aleatory)
Pulse time 1 ms
Release time 10 - 100 ms (aleatory)
How to Change the Pulse Shape
(1) With the Data Selector <2>, click on the symbol . The symbol starts flashing.
(2) Turn the Data Selector <2> to the right by one position to make the pulse shape change and press
it for confirmation.
PHYSIOMED ELEKTROMEDIZIN AG
3 Overview of Current Modes
IONOSON-Basic 29