Poseidon SE7EN+ User Manual Appendix 3 Page 94
WARNING:
Always replace the CO
2
Absorbent cartridge with a new, repacked
absorbent cartridge whenever the oxygen cylinder is re-lled. This
will minimize the risk of CO
2
poisoning!
Intact Breathing Loop Verication
The SE7EN is tested for CO2 duration in two ways with the following with test setup according to
EN14143:2013, a respiratory minute volume of 40 liter/minute in cold water 4° Celsius and a CO2
rate of 1.6 liter/minute.
Scrubber duration standard test result:
A 40 meter static test – 130 min
Note: A static test is a simulated dive performed in a pressure chamber where the depth is
constant, i.e. there is no descent nor ascent in the dive.
Scrubber duration result at non-standard dive prole:
A 100m dive prole with a bottom time of 10 minutes – 200 min
Inspect all breathing-hose connections to ensure that they are properly attached. The attachment
nuts should be hand tight and the nuts should be screwed down ush against the receiver
manifolds in all eight locations (two at the top of the gas processor; four at the shoulder ports;
and two at the mouthpiece). Also at this time, make sure the right counterlung dump valve is
fully closed (turned all the way clockwise). This is important for the pre-dive routine that automati-
cally follows power-up.
CO
2
Absorbent Cartridge - Repackable Scrubber
Guidelines for Different Dive Plans
Follow the procedures described in Chapter 1 for the installation of the Repackable
Scrubber Cartridge, Sofnolime 797. When conducting a repetitive dive, it’s critical to keep
track of the hours of personal use for the cartridge once installed. The absorbent cartridge
must be changed whenever the oxygen cylinder is re-lled. While many people experience a
strong reaction to CO
2
buildup (as would result from diving with a depleted or missing
cartridge) in the form of un-naturally rapid breathing rate, disorientation, and the onset of an
intense headache, some people do not experience these symptoms. Do not risk CO
2
poisoning! Change the cartridge every 2 hours and 10 minutes* of use or whenever the
oxygen cylinder is re-lled, whichever comes rst.
*) Tested static 40m with 797 Sofnolime in cold water 4° Celsius at a breathing rate of 40 liter/
minute RMV and a CO
2
production rate of 1.6 liter/minute.
Mushroom Valve Test
Rubber Mouthpiece Check
Remove the mouth peice at put it in CC position. Breath in and block one side of the hose con-
nections. Here you shall be able to feel the blocking function of mushroom valves on each side.
Also make a visual inspection of the silicone rubber. It should look fresh and not brown/yellow and
dried out.