Poseidon SE7EN+ User Manual Appendix 1 Page 78
T# Time (sec) Description Error Code Solution
29
31
30
Audio Alarm Speaker. This test measures the amount of elec-
trical current consumed by the speaker in the battery (Audio
Alarm), when activated.
Diluent Cylinder Pressure Sensor Validation. This test includes
a series of tests that conrm that power can be supplied to the
diluent cylinder pressure sensor, and that the signal from the
sensor is within limits (regardless of whether the cylinder valve
is turned on).
Oxygen Cylinder Pressure Sensor Validation. This test includes
a series of tests that conrm that power can be supplied to the
oxygen cylinder pressure sensor, and that the signal from the
sensor is within limits (regardless of whether the cylinder valve
is turned on).
4.5
7.5
7.5
11=Current
too low
12=Current
too high
17=Locked on
18=Locked off
19=Defective
14=Locked on
15=Locked off
16=Defective
1) Standard Response; 2) If test failure persists, or the Audio Alarm speaker
does not sound during this test, contact an authorized Poseidon Service
Center for repair.
1) Standard Response; 2) If test failure persists, contact an authorized
Poseidon Service Center for repair.
1) Standard Response; 2) If test failure persists, contact an authorized
Poseidon Service Center for repair.
27 Diluent Calibration Solenoid. This test measures the amount of
electrical current consumed by the diluent calibration solenoid
valve, when activated. It does not check that the solenoid
actually opens and closes (veried during the Positive Pressure
Loop Test, t49).
9
11=Current too low
12=Current too high
87=Voltage too low
88= Voltage
too high
1) Standard Response; 2) If test failure persists, or if no faint “click” sound
from the main electronics module can be heard at the start of this test,
contact an authorized Poseidon Service Center for repair.
26 Oxygen Calibration Solenoid. This test measures the amount
of electrical current consumed by the oxygen calibration
solenoid valve, when activated. It does not check that the
solenoid actually opens and closes (veried during the Positive
Pressure Loop Test, t49).
9 11=Current too low
12=Current too high
87=Voltage too low
88= Voltage
too high
1) Standard Response; 2) If test failure persists, or if no faint “click” sound
from the main electronics module can be heard at the start of this test, contact
an authorized Poseidon Service Center for repair.
32 Bluetooth
7
1) Standard Response; 2) If test failure persists, contact an authorized
Poseidon Service Center for repair.
Note: Bluetooth function is not critical for dive.
---