■ 5 Working with the configuration menu
34
• with “PRODUCTMODE HOLD” this product name is stored,
• with “PRODUCTMODE DELETE” it is deleted after each expression,
• with “PRODUCTMODE COUNT” a counter, which is incremented by
1 after each expression,
• “OPERATOR ttt...” is the alphanumeric input of the operator.
When a peripheral device (for example a printer) is connected, the
balance interface must be configured in the submenu “SET INTER-
FACE” (see Chapter 5.7 “Interface-functions”).
5.5 Calibration functions
To activate the menu see Chapter 4.5 “Activating the two main me-
nus” and Chapter 4.6 “How the menu control operates”.
For the calibration of the balance see Chapter 3.7 “Calibration of the
•SET CALI-
BRATION
MODE OFF Closed
MODE EXTERNAL External
MODE EXT-DEF. External with user-defi-
ned weight (DEF n.nnn g)
MODE INTERNAL with internal weight
MODE AUTO Automatic (AUTOCAL)
DEF. 0.000 g Calibration weight for
EXT.-DEF. mode
AUTOCAL. TIME/TEMP.
AUTOCAL. TIME
AUTOCAL.TEMPERATURE
Autocal. on time/temp.
Autocalibration on time
Autocal. on temperature
AUTOCAL.-TIME 6 h Time for Autocalibration
AUTO-WARNING ON/OFF Shows a message if the
balance recognizes a si-
gnificant temperature
change and therefore a
calibration is recommen-
ded.
REPORT ON/OFF Printout calibration report