5 Working with the configuration menu ■
35
balance” and Chapter 14.2 “Notes on calibration”.
5.6 Weighing mode
To activate the menu see Chapter 4.5 “Activating the two main me-
nus” and Chapter 4.6 “How the menu control operates”.
With the aid of the weighing mode functions, you describe the quality
of the balance location (see Chapter 3.4 “Choice of a suitable locati-
on”).
With the aid of the „AUTO-STANDBY“ function, you can define the
NOTE
The works setting depends on the model of balance. The internal
calibration modes are only available if the self calibration system
(SCS) is installed.
•SET WEIGH-
ING MODE
FLOATINGDISPLAY 0.04
FLOATINGDISPLAY 0.08
FLOATINGDISPLAY 0.16
FLOATINGDISPLAY 0.32
Input lntegration time
(in seconds)
STABILITY LOW
STABILITY MEDIUM
STABILITY HIGH
Setting the Stability
control (instability of
the balance location)
AUTO-STANDBY OFF
AUTO-STANDBY 0.5 MIN
AUTO-STANDBY 1 MIN
AUTO-STANDBY 5 MIN
AUTO-STANDBY 10 MIN
Auto-Standby not-acti-
ve or active after
nn minutes
AUTO-ZERO ON/OFF Automatic zero correc-
tion on/off
QUICK-TARE ON/OFF Quick tare on/off
ZERO-RANGE 0.000 g Maximum weight where
a zero function in per-