14 Further information
67
14 Further information
14.1 Notes on the Weighing mode
14.1.1 Set Weighing mode: FLOATINGDISPLAY
The value set for FLOATINGDISPLAY defines the period, after which
each new measurement is displayed.
For the definition of this period, the quality of the balance-location is
crucial. The stability control must also be suitably chosen.
Recommended values:
14.1.2 Set Weighing mode: Stability control
The value set for the stability control depends on the quality of the
balance location and must be correctly chosen in order to obtain
optimal, reproducible results. Choose:
HIGH STABILITY” at an optimal balance location,
“MEDIUM STABILITY” at a good balance location or
“LOW STABILITY” at a critical balance location
14.1.3 Set Weighing mode: Auto-Standby
The Auto-Standby mode turns off the balance automatically, if:
the balance is tared and has shown “Zero” for at least 5 minutes
the balance has received no remote control command via the inter-
Optimal balance location: FLOATINGDISPLAY 0.04 or
FLOATINGDISPLAY 0.08
Good balance location: FLOATINGDISPLAY 0.16
Critical balance location: FLOATINGDISPLAY 0.32
NOTE
The value of the Floating Display is a function of the stability con-
trol and the balance location. For balance location, see Chapter 3.4
“Choice of a suitable location” and Chapter 5.6 “Weighing mode”.