■ 5 Working with the configuration menu
36
period of non-use before the balance goes over automatically into the
energy-saving mode.
The „ZERO-RANGE“ defines the maximim weigth to performe a zero
function instead of a normal tare when pressing «T» or sending the
remote command to tare. This value can be adjusted to make sure a
small load is saved as tare and printout e.g. on a report as tare value.
5.7 Interface-functions
To activate the menu see Chapter 4.5 “Activating the two main me-
nus” and Chapter 4.6 “How the menu control operates”.
* setting not available for firmware Nxx
** for firmware Nxx factory setting is 600, for other firmware 9600
NOTE
The Auto-Standby function only works with the automatic zero-
correction activated (see Chapter 5.6 “Weighing mode”).
•SET
INTERFACE
BAUDRATE 300
BAUDRATE **600
BAUDRATE 1200
BAUDRATE 2400
BAUDRATE 4800
BAUDRATE **9600
BAUDRATE 19200
Select baud-rate
PARITY 7-EVEN-1STOP
PARITY 7-ODD-1STOP
PARITY 7-NO-2STOP
PARITY 8-NO-1STOP
PARITY *8-EVEN-1STO
PARITY *8-ODD-1STO
Select parity
HANDSHAKE NO
HANDSHAKE XON-XOFF
HANDSHAKE HARDWARE
Enter handshake
function
PC DIRECT MODE*ON/OFF Set PC direct mode