■ 7 Special operating keys
50
7 Special operating keys
7.1 The Tare key «T»
• Effects taring
– Ensure that the balance is in the weighing mode
– Briefly press «T»
– The balance performs a tare operation.
• Effects a calibration
– Ensure that the balance is in the weighing mode
–Keep
«T» pressed until “CALIBRATION” is displayed
–Release «T»
– The balance carries out a calibration in accordance with the set-
tings in the configuration menu (cf. Chapter 5.5 “Calibration
functions”) and reports these by means of a printout.
NOTE
A calibration can be cancelled with «ON/OFF».