9 Error messages and correction of faults
57
9 Error messages and correction of
faults
The balance shows a description of the fault in the info-line.
9.1 Notes on correcting faults
The following table shows faults and their possible causes. If you
cannot clear the fault on the basis of the table, please contact a ser-
vice sngineer.
NOTE
If an error occurs without a description of the error in the info-line,
a service engineer must be called.
Fault
Possible cause
Weight dis-
play does not
light
Balance not switched on
Connection to power adaptor is interrupted
Power supply has failed (interruption to cur-
rent)
The power adaptor is defective
“OL” is shown
in display
The weight range has been exceeded (Observe
information on the maximum weight range)
“UL” is shown
in display
The weight range is below the range of the ba-
lance (Scale pan or scale pan holder missing)
The weight
display
fluctuates
continuously
The draft is too strong at the balance location
The balance support is vibrating or varying
The scale pan is touching a foreign body
The time chosen for FLOATINGDISPLAY is too
short
The material being weighed is absorbing moi-
sture
The material being weighed is being blown
away, is evaporating or is subliming
Strong temperature variations in the material
being weighed