14 Further information
68
face for at least 5 minutes,
the automatic zero correction “Auto-Zero” is activated.
It is possible to start the balance again after it has been switched off
by an automatic Auto-Standby:
Briefly depress any key
Put a weight on the scale
Give a remote control command via the interface
14.1.4 Set Weighing mode: Auto-Zero
If the automatic zero correction “Auto-Zero” is activated, the balance
always gives a stable zero (e.g. even with room temperature fluctua-
tions).
14.2 Notes on calibration
The calibration of the balance is fixed in the configuration menu (see
Chapter 3.7 “Calibration of the balance” and Chapter 5.5 “Calibration
functions”).
Possible types of calibration, depending on the model of balance:
External calibration by means of ICM (Intelligent Calibration Mode)
External calibration with freely selectable weight
Internal calibration
Automatic calibration
14.2.1 External calibration by means of ICM
Depending on the type of balance, calibration weights in steps of
10 g, 50 g, 100 g and 500 g can be used, where the calibration
weight must correspond to the precision of the balance.
For an external calibration by means of ICM, “SET CALIBRATION
NOTE
The calibration can be interrupted at any time by pressing
«ON/OFF».