9 Error messages and correction of faults
58
Tab. 9.1 Possible faults and their removal
Results of
weighing are
clearly incor-
rect
The balance was not correctly tared
The balance is not correctly levelled
The calibration is no longer correct
There are strong temperature variations
There is no
display or only
dashes
The stability control (Balance functions) is set
too sensitively
The time selected for FLOATINGDISPLAY is
unsatisfactory
Configuration
menu cannot
be changed
The password lock is activated in the configu-
ration menu
The display
flashes conti-
nuously du-
ring calibration
The balance location is not quiet enough (Inter-
rupt calibration with «ON/OFF» and relocate
the balance in a better position)
Use of an imprecise calibration weight (only
applies to external calibration)
Fault
Possible cause