AutoCode
®
Blood Glucose Meter
STEP 4: Apply Control Solution
1. Shake the control solution bottle well, then
remove the cap.
2. Squeeze the bottle and discard the rst drop,
then wipe the dispenser tip with clean tissue paper
or cotton.
3. Squeeze the bottle again to
get a second drop onto a clean,
non-absorbent surface or on
your nger. See Figure 18.
4. Bring the tip of the test strip to
touch the drop of solution un
-
til the meter beeps.
32
Figure 18
Caution: To avoid contaminating the control
solution with the content of the test strip, do not
apply the control solution directly onto the test
strip from the bottle. See Figures 19 and 20.