Owner’s Manual
AutoCode
®
Blood Glucose Meter
34
QUESTIONABLE RESULTS:
If the test result falls outside the range printed on the
test strip vial, check the “Troubleshooting Guide” locat-
ed in the “System Troubleshooting” section of the Own-
er’s Manual and repeat the test.
Results that fall outside of the control range may indi
-
cate:
An error in how the test was performed
Control solution bottle not shaken well
Expired or contaminated control solution
Control solution that is too warm or too cold
Test strip deterioration
Meter malfunction
Test strips expired
Caution: Do not use the test strips or control
solution after the expiration date, as the results may
be inaccurate.
Contact Customer Care
If your test strips have been damaged or if you
continue to get out-of-range results, it means that the
system or control solution may not be working properly.
Do NOT use the system to test your blood glucose
level and contact Customer Care at 1.800.243.2636.

Table of Contents

Other manuals for Prodigy Autocode

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Prodigy Autocode and is the answer not in the manual?

Prodigy Autocode Specifications

General IconGeneral
Measurement Range20-600 mg/dL
Sample Size0.6 µL
Test Time6 seconds
CodingNo Coding Required
Battery Type2 x AAA batteries
Automatic Shutoff2 minutes
Data PortUSB
Operating Conditions50-104°F (10-40°C)
Memory CapacityUp to 450 test results

Related product manuals