1.1 Warnings
Warnings:
Only personnel trained and authorized by the manufacturer are permitted to operate
the equipment. Operations of the equipment must be strictly according to the user
manual.
Please read this manual carefully before operating the Boaray 5000C/D Ventilator.
Users must take the related responsibility for any malfunction which results from the
package or product disassembled by the user.
Only those conform to the latest IEC 60601-1 standard accessories and auxiliary
equipment can be connected to the ventilator. If peripheral equipment such as
computer, monitor or humidifier has been connected to the ventilator, the whole
system should meet the IEC 60601-1 standard.
All analog or digital products connected to this system must be certified passing the
specified IEC standards (such as IEC 60601-1 for medical electrical equipment and
IEC 60950 for safety of information technology equipment). All configurations shall
comply with the valid version of IEC 60601-1. The personnel who are responsible for
connecting the optional equipment to the I/O signal port shall be responsible for
medical system configuration and system compliance with IEC 60601-1-1.
Users have the responsibility to carry out the necessary measure to ensure that the
curing environment is in line with the limited regulation of the IEC 60601-1-2 standard.
Operating the equipment beyond the regular limit may cause damages or security
risks to the system. The preventive measures may include (but not limited to) the
following aspects:
More attention should be paid to the relative humidity and conductivity of
clothes to minimize the possibility of the static gathering.
Avoid using the wireless radiate devices (such as the mobile phone) or high
frequency devices near the ventilator.
Due to the possible fire or explosion hazard, all the ignition sources must be away