- To select ventilation mode
- To set ventilation parameters
- Trigger setting: 【Flow Trigger】,【Pressure Trigger】, and 【Close】
- To set backup mode (SPONT mode): 【PCV】,【VCV】,【PRVC】
- Free flow setting: 【Square】,【Ramp】and 【Exit】
Mode and parameters setting
1) In ventilation setting screen, touch or rotate the navigation wheel to【ventilation
mode】 option frame, and select the ventilation mode in the drop-down menu.
2) Select the ventilation parameters that need to set, and rotate the navigation wheel to
adjust the value and push the navigation wheel to confirm setting.
3) Backup mode can be set in SPONT mode. Tou ch 【Backup Mode】 option frame and
select 【PCV】,【VCV】or【PRVC】 in the drop-down menu.
4) Touch 【Free flow】 option frame to select 【Start】or【Exit】.
Note:
The ventilation parameters is corresponding to the current ventilation mode. In other
word, when you select a ventilation mode, only the parameters associated with the
current ventilation mode will display on the screen.
5.8 Monitored parameter screen
Touch the 【measured value】 touch screen icon to enter the measured value screen. The
monitored parameters screen displays measured values and setting parameters. it shown
as below.