B.2 Technical parameters
Parameter
Description
Category
Pneumatic electronically contronlled ventilator
Display
15inch, TFT display, touch screen (optional)
Battery
Lithium batteries, power supply for at least 60 minutes.
Respiration mode
VCV, PCV, SIMV+VCV, SIMV+PCV, SIMV+PRVC, PSV, SPONT,
NPPV(optional), PRVC(optional), DPAP (optional)
Graphics display
Waveform: prssure-time, flow-time,volume-time.
Loop graph(optional):pressure-volume,flow-volume,flow-pressure.
Others(optional): CO
2
, SPO
2
Safe pressure in the
airway system
≤125cmH
2
O.
Data communication
interface
RS232 serial port,
communication interface, LVDS interface
Sound pressure
levels(Normal
operation)
65dB (A)
B.3 Setting parameters
Parameter Description Accurancy
Tidal volume
Range: 202000 mL (D series),
Range: 501500 mL (C series)
Increment: 20100 Ml: 5 mL
1001000 mL: 10 mL
10002000mL: 50 mL
±20mL,
or ±15% of the set
value, whichever is
greater
Respiration frequency
Range: in SIMV mode: 140 bpm
In other modes: 4100 bpm,
increment: 1 bpm
±2bpm or ±10% of the
set value, whichever is
greater
Inspiration time
Range: 0.110.0 s
Increment: 0.1 s
±0.1s
or ±10% of the
set value, whichever is
greater
Breathing hold time
Range: 04 s, increment: 0.1 s
Pressure trigger
Range: -20cmH
2
O-1
Increment: 1 cmH
2
O
<-5 cmH
2
O: ±1 cmH
2
O,
the other is ±20%
Flow trigger
Range: 0.520 L/min
Increment: 0.5 L/min
<-5 L/min : ±1 L/min,
the other is ±20%
PEEP
Range: 050 cmH
2
O
Increment: 1 cmH
2
O
Pressure support
Range: 170 cmH
2
O
Increment: 1 cmH
2
O