2.10.2 Alarm information and priority
Alarm type Message Alarm priority
Physiological alarm
Tidal Volume Low!! Med
Tidal Volume High!! Med
Minute Volume High!! Med
Minute Volume Low!! Med
Frequency High! Low
Frequency Low! Low
FiO
2
Concentration High!! Med
FiO
2
Concentration low!!! High
Airway Pressure High!!! High
Airway Pressure Low!!! High
High Continues Pressure!!! High
Apnea!!! High
PRVC Airway Pressure limit!!! High
Expiratory CO
2
High!! Med
Expiratory CO
2
Low!! Med
SPO
2
low!! Med
System leakage!!! Please check the
breathing tubes and expiratory valve.
High
Technical alarm
O
2
supply pressure low!!! High
Air supply pressure low!!! High
AC Power Failure!!! High
Battery failure!! Med
Low battery capacity!! Med
Battery is exhausted!!! High
O
2
sensor is not connected or failure!!! High
The pipe is failed!!! High
Pressure Sensor is failure!!! High
5V Power Failure!!! High
10V Power Failure!!! High
-12V Power Failure!!! High
12V Power Failure!!! High