85
Alarms content Possible reasons Solutions
to low tidal volume.
The inspiratory valve faiures and leads to
low inspiratory tidal volume.
The pipe line is pluged, and the pressure
protecting leads to low tidal volume.
technicians.
C.3 Low priority alarms
Alarms content Possible reasons Solutions
Frequency high
The breathing frequency is too
high.
Triggering automatically.
Leakage in the pipeline.
Check and take care of the
patient
Check the trigger settings.
Check the connection of the
patient’s pipiline.
Frequency low
The breathing frequency is too
low.
Improper setting of the trigger
sensitivity.
High tidal volume.
Check and take care of the
patient
Check the trigger settings.
Check t
he inspiratory terminate
settings.