Prunus Boaray 5000C - Alarm Silence and Reset

Default Icon
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
different modes determines whether the volume or the pressure control mode will be
selected.
The SIMV mode is combined with the PSV mode, and forms SIMV+VCV+PSV,
SIMV+PCV+PSV and SIMV+PRVC+PSV modes.
In the SIMV mode, the following parameters need to be set:
- P
supp
- Freq
- T
I
- V
T
or P
insp
- FiO
2
- PEEP
- P
TRIG
or F
TRIG
- T
p
- ETS
- P
rate
2.8.6 DPAP mode
DPAP (Dual Positive Airway Pressure). The ventilator forms the high pressure level and
low pressure level according to the pre-set value, and switches by the pre-set frequency
and inspiratory time. Patients can breathe spontaneously at the high pressure and low
pressure level by trigger, and the ventilator provides support ventilation according to the
pre-set pressure.
In the DPAP mode, the following parameters need to be set:
- T
I
- Freq
- ETS
- FiO
2
- P
high
- P
low
- PS
high