79
Parameter Description Accurancy
P
peak
(pressure support) = pressure setting
(Pressure support) + PEEP
±2cmH
2
O or ±10% of
the set value,
whichever is greater
Pressure control
Range: 570 cmH
2
O
Increment: 1 cmH
2
O
P
peak
(pressure control) = pressure setting
(pressure control) + PEEP
High pressure
level(optional)
Range: 570 cmH
2
O
Increment: 1 cmH
2
O
Not suitable for C series, and optional for D series
Low pressure
level(optional)
Range: 050 cmH
2
O
Increment: 1 cmH
2
O
Not suitable for C series, and optional for D series
High pressure
support(optional)
Range: 170 cmH
2
O
Increment: 1 cmH
2
O
Low pressure
support(optional)
Range: 170 cmH
2
O
Increment: 1 cmH
2
O
Oxygen concentration
Range: 21100%
Increment: 1%
±6% of the set value
Apnea time
Range: 1060 sec
Increment: 5 sec
±1 s or ±10% of the set
value
Respiration sensitivity
Range: 1060%
Increment: 10%
P
rate
Range: 15
Increment: 1
B.4 Monitoring parameters
Parameter Description
Inspiration tidal volume
Range: 03000 mL, Resolution: 1 mL
Expiration tidal volume
Range: 03000 mL, Resolution: 1 mL
Minute ventilation volume
Range: 099 L, Resolution: 0.1 L
S
pontaneous minute ventilation
volume
Range: 099 L, Resolution: 0.1 L
Total respiration frequency
Range: 0100bpm, Resolution: 1 bpm
S
pontaneous respiration
requency
Range: 099 bpm, Resolution: 1 bpm
Inspiration and expiration rate
Range: 4:11:9, Resolution: 0.1
PEEP
Range: 050 cmH
2
O, Resolution: 1 cmH
2
O
Airway peek pressure
Range: 0100 cmH
2
O, Resolution: 1 cmH
2
O
Average airway pressure
Range: 0100 cmH
2
O, Resolution: 1 cmH
2
O

Table of Contents